1. The purpose and characteristics of the testing machine
The JBW-300B microcomputer-controlled pendulum impact testing machine produced by our company is a new generation of high-tech testing equipment developed according to the national standard GB/T3808-2010 "Inspection of Pendulum Impact Testing Machines". Charpy Notch Impact Test Method" Impact test on metallic materials. This machine adopts the method of replacing the pendulum to achieve two-level impact energy of 300J and 150J. The cross-section of the sample used is 10×10mm. It is mainly used to test ferrous metals with high impact toughness, especially steel and its alloys.
The test principle of this product is to calculate the difference between the potential energy before the impact of the pendulum and the remaining potential energy after the impact to obtain the absorbed energy of the tested sample. The operation of the equipment can be controlled by a manual control box or a computer automatically. The operation is simple and efficient. The remaining energy after the pendulum breaks the sample can be used to automatically raise the pendulum, and the test results can be saved and printed. In the continuous impact test of the sample, its superiority can be more manifested.
2. Main technical specifications
1) Impact energy: 300J,150J
2) Scale range and division value:
Energy range | 0~300J | 0~150J |
Division value per cell | 2J | 1J |
Pendulum impact energy | 300J | 150J |
Pendulum moment | 160.7695N.m | 80.3848N.m |
4) Pendulum advance angle: 150º
5) Distance from pendulum center to impact point (test center): 750mm
6) Impact velocity: 5.2m/s
7) Span of sample support: 40mm
8) The arc radius at the end of the specimen support: 1~1.5mm
9) The inclination angle of the supporting surface of the specimen support: 0º
10) The arc radius of the impact knife: 2~2.5mm
11) Including angle of impact knife: 30º
12) Thickness of impact knife: 16mm
13) Specimen specification: 10×10×55mm
14) Weight of testing machine: about 450kg
15) External dimensions of the testing machine: 2124×600×1340mm
16) Power supply: three-phase four-wire system 50Hz 380V
17) Motor power: 380V 180W
18) The minimum resolution of computer display: 0.1J
19) Working conditions:
a) The room temperature is within the range of 10-35ºC;
b) The relative humidity is not more than 85%;
c) In an environment without corrosive media and strong magnetic field interference;
d) Installed on a concrete foundation with a thickness of not less than 300mm or fixed on a foundation greater than 880 kg;
e) Adjust the level of the installation reference surface on the machine base to within 0.5/1000.
3. Structure introduction
The JBW-300A screen display impact testing machine is composed of the following parts: (1) Body; (2) Swing mechanism; (3) Hang and release swing mechanism; (4) Signal measuring device; (5) Dial display device; (6) Pendulum; (7) Safety device; (8) Computer; (9) Electrical part (with special PLC).
JBW-300B microcomputer controlled pendulum impact testing machine Certificate of Qualification |
||||
Factory number:CY20201024 | ||||
Serial number | Inspection items | Content | Tolerance | Conclusion |
1 | Appearance quality | |||
2 | Installation quality | |||
3 | Base level | ≤2/1000 | ||
4 | When the pendulum is vertical, the distance between the impact edge and the supporting surface of the anvil is 10mm | ±0.5mm | ||
5 | Parallelism and perpendicularity of anvil and support | The two supporting surfaces of the anvil should be parallel | ≤0.1mm | |
The angle between the plane where the two supporting surfaces of the support are located and the plane where the two supporting surfaces of the anvil are located | 90°±0.1° | |||
The two supporting surfaces of the support should be parallel | ≤0.1mm | |||
6 | Parallelism between the side of the pendulum and the swing plane |
Big Pendulum | 1/1000 | |
Small Pendulum | ||||
The perpendicularity between the side of the pendulum and the specimen support | Big Pendulum | 90°±0.1° | ||
Small Pendulum | ||||
7 | The impact blade should pass through the center of the support span | ≤0.5mm | ||
8 | Axial clearance of pendulum shaft | ≤0.25 mm | ||
Radial clearance of pendulum shaft | ≤0.08 mm | |||
9 | Pointer back to zero (visual inspection) | Big Pendulum | ±0.3J | |
Small Pendulum | ±0.15J | |||
10 | Energy loss |
Big Pendulum | ≤1.5J | |
Small Pendulum | ≤0.75J | |||
11 | Distance from the axis of the pendulum shaft to the center of strike | 746.25mm | ±3.75mm | |
12 | Parallelism between the axis of the pendulum shaft and the base plane of the base | ≤2/1000 | ||
13 | Anvil span | 40+0.2 0mm | ||
14 | Impact blade angle | 30°±1° | ||
15 | Pendulum torque error | Big Pendulum:160.7695N.m | ±0.4823N.m | |
Small Pendulum:80.3848 N.m | ±0.2412N.m | |||
Conclusion: After the above-mentioned inspection, this testing machine meets the technical requirements of the national standard GB/T229-2007 "Inspection of Pendulum Impact Testing Machine". An impact test was carried out, and the testing machine worked normally, and all precisions were maintained well. The accessories and technical documents are complete. Confirm that it is qualified and allowed to leave the factory. |
JBW-300B
Packing list
Serial number | Name | Quantity |
1 | Host | 1 set |
2 | 150 joule pendulum | 1 set |
3 | 300 joule pendulum | 1 set |
4 | Span aligning plate L=40 | 1 item |
5 | Sample centering hammer L=40 | 1 item |
6 | Allen key S=12 | 1 item |
7 | Remover (for hammer replacement) | 1 set |
8 | Anchor bolt | 4 sets |
9 | Inclined iron | 4 pieces |
10 | Computer (Lenovo brand) | 1 set |
11 | HP printer (inkjet) | 1 set |
12 | Dedicated PLC | 1 set |
13 | Instruction manual | 1 serving |
14 | Certificate of Compliance | |
15 | Packing list | |
16 | Protective net | 1 set |
17 | Install the protective net bolts | 8 sets |
18 | Hand control box | 1 set |
FAQ